|
PHOTOVOLT TECHNICAL NOTE
Procedure For The Analysis Of Water In An Aquatest Vaporizer
Using Capillary Tubes
Weigh a single capillary tube on the pan of a four-place balance
(or preferably on a microbalance). Immerse in water a few millimeters
to draw a small amount of water into the tube. Carefully wipe off
the outside of the tube, replace on the balance pan and re-weigh.
Determine amount of water by difference. Place tube in an aluminum
or glass sample boat.
Insert boat into vaporizer tube, cap oven, push boat to oven entrance,
where it will stop. Press START; the PURGE cycle will begin. When
the PURGE cycle is over, the display will show ADD SAMPLE. Press START
immediately. The test will continue automatically until you are prompted
for mass weight, etc.
PROCEDURE
Temperature
At the temperature of the intended analysis, typically 150șC to 300șC.
Higher temperatures give shorter analysis time. Time will be longer
in glass boats than in aluminum boats.
Purge (min) 1
Preheat 0
Cool Time 0
When display reads ADD SAMPLE, press START immediately.
CAPILLARY TUBES
Clinical type, untreated (no heparine or EDTA).
Melting Point Tubes May have to be broken into smaller pieces (discard
sealed end).
Micro-Pipettes Any size-40 microliters satisfactory.
Tubes should be capable of containing 1 to 10 milligrams of water.
This quantity of water should result in 1000 to 10,000 micrograms
on the display. A deviation of ± 5% is considered satisfactory. This
is the same deviation we allow for direct injection of water into
the titration vessel for quality control procedures.
1996-09-12
Revised 1999-02-12
|