Aquatest Karl Fischer Titrator
For Coulometric Moisture Determinations
The
determination of water is one of the most important and most widely
practiced analyses in industry. The field of application is so large
that it is the subject of a 3-volume monograph Aquametry.(1) Of all
the techniques the Karl Fischer titration(2) is the most useful and
generally applied. It relies on the specificity to water of the reagent
devised by Fischer. The latter contains pyridine, sulfur dioxide,
and iodine in an organic solvent. It reacts with water quantitatively:
C5 H5
N· 12 + C5 H5
N·SO2 + C5 H5
N + H2 O --->2C5 H5
N· HI + C5 H5 N· SO3
There is a secondary reaction with the methanol (solvent):
C5 H5
N·SO3 + CH3 OH--->C5
H5 NH· SO4 CH3
It is noted that although this reaction is specific and quantitative,
some difficulty may be caused by any admixed impurities which can
react with the solvent to form water, interfere by redox reaction,
or which bind iodine or react chemically with the other constituents.
Usually these side reactions are much slower than the reaction with
water, and although they might be discerned, such reactions cannot
be eliminated. Altogether, the Karl Fischer titration technique is
successful in most problems and its use has shown a steady growth
over the past 40 years.
Various improvements of the original technique are also of long standing.
Karl Fischer himself suggested that an electrometric endpoint detection
could be used instead of the visual observation of the disappearance
of the dark color of iodine. Almy and Griffin(3) used a potentiometric
endpoint successfully. However, an even sharper endpoint detection
was introduced by Wernimont and Hopkinson(4) who employed the "dead-stop"
method of polarized sensing electrodes.(5) The various commercial
instruments are all based on this method, by now considered classic.
However, in 1959, Meyer and Boyd demonstrated the possibility of coulometrically
generating iodine in Karl Fischer titrations.(6) This was a novel
concept with great potential advantages. If the reagent mix contains
iodide together with pyridine, sulfur dioxide and a solvent, the necessary
iodine to complete the Karl Fischer reaction can be generated by passing
electric current through the titration chamber. When the geometry
of the cell and other conditions are properly chosen, 100% current
efficiency can be maintained. When this is the case, according to
the basic laws of coulmetry, 96,500 coulombs = 1 chemical equivalent
(Faraday number). A Karl Fischer coulometer is then an absolute instrument,
and the analysis requires no calibration or standardizaton. The coulometer
can be governed by the classic dead-stop electrode. As the latter
governs the reaction back to the original set-point, sample after
sample can be titrated in the same solution. This system then can
be essentially all-electronic, obviating the need for handling liquids,
burettes, etc.; this is of particular advantage in this case because
the Karl Fischer reagent is noxious and somewhat toxic. Additionally,
an electronic system lends itself to automation. These advantages
were recognized and a number of coulometric Karl Fischer titrators
were employed in research.(7) The first demonstration of such a system
in routine analysis was by Shaw and Goode.(8) Encouraged by this background
information the first commercial instrument, the AQUATEST,(9) was
produced by Photovolt. It had a typical accuracy of 1% and sensitivity
of about 20 micrograms of water. What made it particularly attractive
as an industrial quality control instrument was that it read directly
in micrograms of water and that the only control was a "Start" switch
to initiate the titration. One vexing problem was keeping the reagent
free from deterioration by ambient humidity; thus this model was designed
with a completely closed titration vessel. However, whenever positive
or negative pressure developed, the liquid was pumped from the cathode
to the anode, or vice versa. It was then found that the vessel solution,
which lacks iodine, is not nearly as hygroscopic as the usual Karl
Fischer reagent. Consequently, a new version, the AQUATEST II, was
built with a pressure-equalizing vent. The cathode chamber and configuration
was also improved and the instrument exhibited superior performance.(10)
The AQUATEST II enjoys widespread use and a number of papers have
been published on its performance in the field.(11) It is also accepted
in an official test method.(12)
The AQUATEST IV was a newer version of these instruments. It is the
result of a further study(13) in which the limiting factors were scrutinized.
Changes include an asymmetric sensing electrode, a cover fastened
by a clamp which reduces still further the seepage of ambient humidity,
and a stepping motor that provides more reliable performance. Together,
these changes tend to reduce electrical noise.
The AQUATEST IV incorporated a microprocessor which offers the advantages
pointed out above. The most important is, perhaps, that it serves
to distinguish between the titration of water in the sample and water
which is slowly generated by parasitic reactions, ambient humidity
which seeps in, or any other effects which are oxidizing (or reducing)
in character. All these effects, conveniently called "drift", are
determined subsequent to the titration proper and the result of the
titration is automatically corrected. The microprocessor also performs
the arithmetic operations. There are no controls other than the keyboard,
which incorporates all the functions.
The latest version of the AQUATEST is the AQ2010.
The operation is based on the principles of all AQUATEST instruments.
When the instrument is started and a sample is added to the vessel
solution, a voltage arises across the polarized sensing electrode
which indicates a "wet" condition. This triggers the coulometer and
a constant current flow from the anode through the frit, which separates
the vessel solution (anolyte) from the generator solution (catholyte),
to the cathode. In consequence, iodine is developed at the anode by
oxidation of iodide. The iodine completes the Karl Fischer reagent
and is mixed by stirring throughout the vessel. When all the water
has reacted, the voltage at the sensing electrode drops. This signals
the coulometer to stop, and the electrical charge integrated during
the titration process is stored in memory. During the subsequent time
period of approximately one-minute, data is taken to establish current
requirement, both positive and negative, to maintain the solution
at equivalence. These data are then used to correct the initial titration
value to reflect the net value due only to the water content of the
sample without the background "drift".
1. Mitchell, J. Jr.; and Smith, D.M., Aquametry John
Wiley & Sons, New York, Second Edition 1977
2. Fischer, K., Angew, Chem. 48, 394 (1935)
3. Almy, E.G., Griffin, W.E., and Wilcox, C.S., Ind. Eng. Chem., Anal.
Ed. 12, 392 (1940)
4. Wernimont, G., and Hopkinson F.J., ibid. 15, 272 (1943)
5. Foulk, C.W., and Bawden, A.T., J.Am. Chem. Soc. 48, 2045 (1926)
6. Meyer, A.S. Jr., and Boyd, C.M., Anal. Chem. 31, 215 (1959)
7. Swensen, R.F., and Keyworth, D.A., ibid 35, 863 (1963) also Pribyl,
M., and Slovak, Z., Mikrochim. Acta, 6, 109 (1964), Lindbeck, M.R.,
and Freund, H., Anal. Chem. 37, 1647 (1965), Rechnitz, G.A., and Srinivasan,
K.Z. Anal. Chem. 210, 9 (1965), Bizot, J., Bull. Soc. Chim. 1967,
Vol 1, 151, Karlsson, R., and Karrman, K.J., Talanta, 18 459 (1971)
8. Shaw, R.G., and Goode, J.V., private communication
9. Seltzer, D.M., and Levy, G.B., Am. Lab. 1, (9), 30 (1969)
10. Levy, G.B., and Seltzer, D.M., Recent Analytical Developments
in the Petroleum Industry, Hodges, D.R., Ed. Chapter XV, Applied Science
Publishers, Essex England (1974)
11. Johnson, C.R., Power, March 1971 also Gedener, T.J., and Frey,
R. Amer.Lab 7 (3) 47, (10) 43 (1975), 8 (7) 41 (1976), Draper, R.,
Bakers Dig., June 1977, p. 26
12. ANSI/ASTM D1533-77 13. Lindblom, K.A., et al. Paper #260 Pittsburgh
Conference on Analytical Chemistry and Applied Spectroscopy, February
1978.
13. Lindblom, K.A., et al. Paper #260 Pittsburgh Conference on Analytical
Chemistry and Applied Spectroscopy, February 1978.
|